
- Notch Filter Defined: A notch filter (also known as a bandstop filter or reject filter) is defined as a device that blocks certain frequencies while allowing others to pass.
- Circuit Overview: The notch filter circuit combines low-pass and high-pass filters in a configuration that blocks specific frequency bands.
- Design Principles: Designing a notch filter involves selecting appropriate values for resistors, capacitors, and inductors to achieve the desired frequency rejection.
- Transfer Function Explained: The transfer function of a notch filter describes how it attenuates certain frequencies while maintaining others, crucial for understanding filter behavior.
- Practical Applications: Notch filters are essential in electronics to remove interference, such as the 50/60 Hz noise in medical ECG readings, enhancing signal clarity.
What is a Notch Filter (Bandstop Filter)?
A notch filter (also known as a bandstop filter or reject filter) blocks a chosen frequency band and passes the frequencies above and below that band. A true notch is a narrow stop. A wider stop is usually called a band-reject or band-elimination filter.
A notch is a band-stop filter with a narrow stopband and two passbands. As in the band-pass case, a band-reject filter can be wide or narrow.
A wide stop is a band-reject filter. A narrow stop is a notch. The magnitude shape is the complement of a band-pass: pass, stop, then pass again.
Example: a stop from 100 MHz to 200 MHz passes DC to 100 MHz and everything above 200 MHz. It rejects 100 MHz to 200 MHz. That span is a wide band-reject, not a sharp audio or mains notch.
In symbols, the filter passes from DC up to the lower cut-off frequency(fL) and again above the upper cut-off (fH). It rejects the band BW = fH-fL.
A common signal processing techniques job is to pull a narrowband interferer out of noise. A notch can remove that tone while leaving the rest of the band energy in place.
Notch Filter Circuit
One wide band-stop build puts a low-pass and a high-pass in parallel, then sums them. A cascade of the same pair would be a band-pass. The figure below is that parallel idea. A twin-T notch is a different, narrower RC network.

The upper arm is a passive RC low pass filter T: two resistors (R1 and R2) and a capacitor (C1). That path passes frequencies below its own cutoff. On a wide band-stop that cutoff sits at the lower stop edge, not at fH.
The lower arm is a passive RC high-pass filter T: capacitors (C2 and C3) and resistor (R3). That path passes frequencies above its cutoff. On a wide band-stop that cutoff sits at the upper stop edge, not at fL. The two T networks together are often called a Twin-T. A classic Twin-T notch is a matched R-C-R and C-R-C pair that rejects one frequency, not a wide LPF-plus-HPF stop.
A typical block view and a notch frequency response are shown below.


Notch Filter Types
Common builds are listed below.
Active Notch Filter
An active notch can be a low-pass and a high-pass in parallel, then an op-amp summer, as in the figure below.

That active circuit has three parts: an active low pass filter (LPF) in parallel with an active high-pass filter (HPF), then a summing amplifier or a Op-Amp adder. The summer adds the two paths. The figure above is that arrangement.
The low-pass cutoff (lower stop edge) is
![]()
The high-pass cutoff (upper stop edge) is
![]()
Passive Notch Filter
A passive notch uses only resistors, inductors and capacitors. There is no op-amp gain stage.
The RC figure below is a passive low-pass T in parallel with a passive high-pass T.

The upper arm is the low-pass T. The lower arm is the high-pass T.
Optical Notch Filter
An optical notch filter blocks a chosen wavelength band and transmits the rest. Rejection is often stated as optical density.
A grating or prism can spread light by wavelength. An optical notch then blocks one of those bands, for example a laser line and passes the others.
Real coatings also lose some light at other wavelengths. A well-made optical notch still passes the wanted bands with only small extra loss.
They are used in laser-based Raman spectroscopy, laser fluorescence and other life-science optics.
RLC Notch Filter
An RLC notch uses the three passive elements R, L and C. One series-LC shunt form is shown below.

In that form the shunt is series L and C, and the output is taken across that series pair. At series resonance the shunt impedance is a minimum, so that frequency is rejected. The equivalent circuit is shown below.

The series L-C impedance is
![]()
The net impedance including R is
![]()
From KVL (Kirchhoff’s Voltage Law) on that loop, the stored voltage ratio is

At series resonance the resonant frequency is
![]()
The series Q-factor is
![]()
Where
is the resonant angular frequency, 2π fr. The 3 dB bandwidth is
, not ωr itself.
Then
is the lower cut-off, and
is the upper cut-off. Those arithmetic-mean edges are a good fit when the stop is narrow. The curve below is the magnitude shape.

Another RLC notch arrangement is shown below.

RF Notch Filter
An RF notch is a narrow stop used, for example, to reject a carrier while measuring amplifier distortion products.
An LC circuit wave-trap can reject one strong interferer so a nearby receiver can still hear weaker stations. That is the old broadcast “wave trap” use.
Strong FM broadcast carriers can also overload an SDR. An FM-band notch ahead of the SDR USB dongle is a common fix.

Butterworth Notch Filter
A Butterworth notch is a Butterworth prototype mapped to a stopband. Butterworth means a maximally flat passband, not a flat stopband.
An ideal notch would reject the stop and pass everything else with no extra distortion. A simple first-order trap is often too shallow when the signal-to-noise ratio is already low. A higher-order accuracy-minded Butterworth notch can make a deeper, better-shaped stop.
On an ECG (electrocardiogram) the usual extras are mains hum, instrument noise, external magnetic field interference, motion artefact and breathing. Each sits in a different frequency range.
Some of that noise sits inside the ECG band, so a blunt filter also hurts the wanted trace. A 4th order Butterworth notch around 50 Hz (or 60 Hz) is a published way to cut mains hum. The “3 dB stopband” wording is kept as the page’s own spec; check the paper or datasheet for the exact edges.
Notch Filter Design Example
Design an RLC notch with cut-offs 23 kHz and 25 kHz. Take Inductance L = 45 mH.
Given: fL=23 kHz, fH=25 kHz, L=45 mH = 0.045 H
- Bandwidth (BW) =

- Resonant frequency

- At resonance,

- Quality factor (Q) =

- Now,

The stored design values are
R=565.2 ohms , L=45 mH , C=977 pF. Those figures match 2π√(LC) ≈ 24 kHz and R = 2π fr L / Q to the stored rounding.
The corresponding RLC notch is shown below.

Notch Filter Transfer Function
A stored second-order notch transfer function is

where
is the zero (notch) angular frequency
is the pole angular frequency
Q is the selectivity.
where BW is the 3 dB stop bandwidth.
The same H(s) gives a standard notch, a low-pass notch or a high-pass notch. Which one you get depends on
versus
.
- If
, the filter is a standard notch. - If
<
, the filter is a low-pass notch (more high-frequency gain). - If
>
, the filter is a high-pass notch (more low-frequency gain).
Notch Filter Applications
Where notch filters are used:
- Communications, instruments and biomedical front-ends use notches to cut 50 Hz or 60 Hz mains.
- Electronics and radio use a band-stop to reject one band and pass the rest with little extra loss.
- Switching AC & DC motor drives, converters and inverters inject tones at harmonics of the line. A notch can remove those tones so a measurement is not swamped.
- In image and signal processing a notch can cut a single interfering tone.
- In audio it is used to cut hum or a whistle.
- Telephone and DSL plant use stop filters on known interferers. Note that DSL is the Digital Subscriber Line that is used to transmit digital information over telephone lines.
- Guitar, bass and PA amps use a notch to cut a plug-in hum or a room whistle. Note that PA (Public Address) systems are an electronic system comprising of microphones, amplifiers, loudspeakers, and other musical equipment.
- In ECG work a 50 Hz or 60 Hz notch cuts mains. It does not remove the DC baseline. A separate high-pass does that.