- Chopper Amplifier Definition: A chopper amplifier is defined as a device that samples and converts DC input signals to AC signals, amplifies them, and converts them back to DC to reduce drift.
- Types of Chopper Amplifiers: There are two types of chopper amplifiers: mechanical and non-mechanical.
- Mechanical Chopper Amplifier: This type uses an electromagnetically operated switch to convert and amplify signals, then converts them back to DC.
- Non-Mechanical Chopper Amplifier: This type uses photodiodes or photoconductors to modulate and demodulate signals, creating a square wave pulse that is amplified and filtered.
- Differential Chopper Amplifier: Used for EEG measurements, it includes a transformer and chopper vibrator to convert and amplify signals, producing a stable DC output.
When recording biopotentials, noise and drift are two problems. Noise can come from the recording device and from the patient when they move. Drift is a shift in baseline created due to various thermal effects. A DC amplifier can show a shift or sudden peak in the output when the input is zero. A chopper amplifier is used in biomedical instrumentation to reduce that DC drift. The name chop means to sample the data. The amplifier circuit samples the analog signal. So it is known as chopper amplifier.
The general diagram of a chopper amplifier includes three main parts. First, the chopper block converts the DC input signal to an AC signal. Then, the AC amplifier block amplifies this AC signal.
Next, in the demodulator rectifier block, an amplified chopped AC signal is converted to an amplified DC signal.
Chopper amplifiers are classified into two types: mechanical and non-mechanical. The chopper converts DC or low-frequency signals to a higher chopping frequency. The AC amplifier then amplifies those chopped signals, which are demodulated and filtered to produce low-frequency or DC signals.
Mechanical Chopper Amplifier

From the figure, chopper S1 acts as an electromagnetically operated switch or relay. ‘A’ is the AC amplifier that has an input terminal and a ground terminal. ‘Q’ acts as reference term. The chopper acts as a switch, so it connects the amplifier input terminal alternatively to reference term Q. Consider a condition in which chopper S1 is closed. At this position, the amplifier input terminal connects to Q1. The entire circuit is short-circuited, so input voltage is zero. Now, let us consider the reverse operation when chopper S1 is open. The AC amplifier starts receiving the signal from P terminal. Finally, the amplifier input has an alternating voltage that varies between zero and input voltage. At this stage, conversion of DC signal to square wave pulse occurs with amplification. Diode ‘D’ rectifies the chopped signal.
After rectification, the rectified signal is filtered and amplified. At the output terminal M and N, the amplified DC output signal occurs. Chopping or sampling rate determines the chopper response time.
Non-Mechanical Chopper Amplifier

Unlike the mechanical type, a non-mechanical chopper uses photodiodes or photoconductors for modulation and demodulation. When no light hits the photodiode, no current flows. When light falls on the photosensor, its resistance decreases, allowing current to flow. This process is similar to a switch.
From the figure, an oscillator has two neon bulbs, which operate on half cycles of oscillation. PC1, PC2, PC3 and PC4 are photodiodes. Neon lamp 1 flashes light on PC1 and PC2. Neon lamp 2 flashes light on PC3 and PC4. When light falls on PC1, its resistance value reduces making the capacitor to charge. Light falls on PC3 making the input to flow through it when there is no light on PC1. Therefore, the light incidence on PC1 and PC3 takes place alternatively to generate a square wave pulse across the output capacitor. The generated square wave pulse is the input for the AC amplifier. The amplifier output is an amplified square wave pulse. The other two photodiodes PC2 and PC4 are in the output circuit. Those diodes recover the DC signal and charge the capacitor toward the peak of the output voltage. At the final stage, a low pass filter removes the leftover chopping ripple. The output is an amplified DC signal.
Differential Chopper Amplifier

A type of chopper once used for EEG measurement is a differential chopper. It has a transformer. A chopper vibrator connects the input of the transformer. The center tap of the transformer acts as one of the terminals for the input connector. The chopper switch acts as another terminal. An AC coupled amplifier provides the gain. The output from this amplifier goes to a filter and demodulator block. Modern EEG front ends more often use an instrumentation amplifier. The vintage chain still yields an amplified DC output signal.